4
Analyze the starting and ending digit of the data read from the magnetic card.
The analyzing procedure is explained using the example of patient data below.
Patient ID
: 1234567890
Patient Name : FUKUDA DENSHI
Date of Birth
: Jan. 1, 1950
Sex :
Male
1) Open the first page of the magnetic card reader setup screen, and scan the magnetic card.
At the first and second row, the data read from the card will be displayed in hexadecimal.
At the third row, the characters converted from the data will be displayed.
2) From the displayed result, specify the data position.
[Patient ID]
The patient ID, “1234567890” is displayed at 1st to 10th digit.
[Patient Name]
The patient name, “(FUKUDA) (DENSHI)” is displayed at 11th
to 23th digit. Depending on the length of patient name, up to
59th digit may be used. “( )” should be also included in the
digit range.
Data read from
the card
(displayed in
code)
Digit Number
8
-
34