62
2
Select the scale for F-V loop.
The scale will change as the Scale key is pressed.
・
Vertical Axis (Flow)
Select from
±
5 /
±
10 /
±
20 /
±
50 /
±
180 (L/min).
・
Horizontal Axis (Volume)
Select from 250 / 500 / 750 / 1000 (mL).
3
Freeze the loop drawing.
Pressing the Freeze key will freeze the F-V loop. Pressing
the key again will resume the waveform trace.
4
Program the reference loop.
The control loop can be stored to see the change in F-V loop.
Pressing the Store key will store the displayed F-V loop as
the control loop. Pressing the Review key will display the
stored control loop.
The control loop 1 will be displayed in yellow, and control loop 2
will be displayed in green.
5
Print the F-V loop.
The currently displayed F-V loop will be printed.
N O T E
The F-V loop cannot be printed on the central monitor recorder.