15
Alarm Setup
Item
Selection
Default
Backup
HR
PR-SpO
2
PR-IBP
ON, OFF 20 to 300bpm
Adult: ON 50 to OFF
Child: ON 60 to OFF
Neonate: ON 70 to OFF
{
/
U
ASYSTOLE
ON 3 to 10sec.
ON 5sec.
VF ON,
OFF
ON
VT ON,
OFF
ON
SLOW VT
ON, OFF
OFF
RUN
ON, OFF 2 to 8 beats
OFF 3 beats
COUPLET ON,
OFF
OFF
PAUSE
ON, OFF 1.5 to 5sec.
OFF 3.0sec.
BIGEMINY ON,
OFF
OFF
TRIGEMINY ON,
OFF
OFF
FREQUENT
ON, OFF
1 to 50 beats/min.
OFF, 10beats/min.
TACHY ON,
OFF
OFF
BRADY ON,
OFF
OFF
{
/
U
ST(
Ι
), ST(
ΙΙ
), ST(
ΙΙΙ
),
ST(aVR), ST(aVL),
ST(aVF), ST(V) (mm)
ST All Alarm ON, OFF
Indiv. Alarm ON, OFF
−
20mm to +20mm
ST All Alarm OFF
Indiv. Alarm OFF
OFF to OFF
{
/
U
ST(
Ι
), ST(
ΙΙ
), ST(
ΙΙΙ
),
ST(aVR), ST(aVL),
ST(aVF), ST(V) (mV)
ST All Alarm ON, OFF
Indiv. Alarm ON, OFF
−
2.0mV to +2.0mV
ST All Alarm OFF
Indiv. Alarm OFF
OFF to OFF
{
/
U
BP1
ON, OFF 0 to 300mmHg,
0 to 40kPa
ON
S
80 to 200mmHg
10 to 27kPa
D
OFF to OFF
M
OFF to OFF
{
/
U
BP2 to BP6
ON, OFF 0 to 300mmHg,
0 to 40kPa
OFF
S
OFF to OFF
D
OFF to OFF
M
OFF to OFF
{
/
U
CVP (mmHg)
ON, OFF 0 to 300mmHg
0 to 40kPa
OFF
S
OFF to OFF
D
OFF to OFF
M
OFF to OFF
CVP (cmH
2
O)
ON, OFF 0 to 40 cmH
2
O
OFF
S
OFF to OFF
D
OFF to OFF
M
OFF to OFF
{
/
U
RR (RR-IMP, RR-CO
2
,
RR-VENT) (Adult)
ON, OFF 5 to 150Bpm
ON 5 to 30
{
/
U
RR (RR-IMP, RR-CO
2
,
RR-VENT) (Child,
Neonate)
ON, OFF 2 to 150Bpm
ON 5 to 30
{
/
U
APNEA
ON, OFF 10 to 60sec.
ON 60 sec.
{
/
U
SpO
2
ON, OFF 70 to 100%
Adult: ON 90 to OFF
Child: ON 90 to OFF
Neonate: ON 85 to OFF
{
/
U
11
Setup Item
Summary of Contents for DynaScope 7000 Series
Page 38: ...xxxvi Blank Page ...
Page 126: ...4 78 Blank Page ...
Page 148: ...5 22 Blank Page ...
Page 259: ...7 15 The Description of the Display Scale Real Time Scale Scale 7 Table Trend ...
Page 322: ...7 78 Blank Page ...
Page 428: ...10 32 Blank Page ...
Page 466: ...11 38 Blank Page ...
Page 474: ...12 8 Blank Page ...