6
-
62
●
Start the Measurement Early
If 60M or 120M is selected for NIBP measurement interval, the measurement can be started 5
minutes before the starting time.
When outputting the measurement data to external equipment (ex. personal computer) using the PC
communication function, some error may be generated to NIBP measurement time depending on the
input interval of the external equipment. As the DS-7000 outputs the measurement data upon
completion of NIBP measurement, if the external equipment inputs the measurement in 60 minutes
interval, 60 minutes difference will be generated. By starting the measurement 5 minutes earlier, this
difference can be minimized.
1
Display the second page of the NIBP setup menu to set the “Start 5min. early”.
ON will start the measurement 5 minutes before.
●
NIBP Data Erase Time
The NIBP data can be erased after preprogrammed duration (60min/120min).
1
Display the second page of the NIBP setup menu to set the “NIBP Data Erase Time”.
Select the duration from 60min. or 120min. .
When the selected duration elapses, the NIBP data will
be erased.