69
2
Press the Operation Room ID key to access the display for the other monitor.
ECG waveform and numeric data for the selected monitor will be
displayed.
If an alarm is generated for this bed, the physiological alarm /
arrhythmia alarm message will be displayed.
By pressing the Alarm Silence key on the other monitor
display, the alarm sound for the displayed monitor can be
silenced.
Other Monitor Alarm Setup
From the 48 monitors, the monitors to generate an Other Monitor Alarm can be selected and ON/OFF
of Other Monitor Alarm display can be set.
1
Select the monitors to generate the “Other Monitor Alarm”.
Press the Set Alarm key to display the “Other Monitor
Alarm” setup menu.
Select the monitors to generate the “Other Monitor Alarm”.
The key LED for the selected monitors will be lighted.
2
Select ON for the other monitor alarm.
Press the Prev. Disp. key to display the other monitor
selection menu.
Selecting ON will generate an “Other Monitor Alarm”
when an alarm generates at the other monitor.
Selecting OFF will not generate an “Other Monitor Alarm”.
7
Other Mo
nitor Data