Cause 2 : The monitor is set to night mode.
Solution : Cancel the night mode.
The system does not start although the power switch is turned ON.
Cause 1 : The power cable is not connected.
Solution : Turn off the power and connect the power cable.
Cause 2 : Incorrect CF card is inserted.
Solution : Remove the CF card, and turn OFF the power. Turn OFF the DIP switch No. 8 and turn
ON the power again.
The clock is often delayed.
Cause
: The battery for the backup memory is depleted. Check if the time is delayed when the
power is turned off.
Solution : The battery needs to be replaced. Contact our service representative.
The “Set DS-7000 Rotary SW0” message is displayed.
Cause
: The rotary switch is not set to “0 (zero)”.
Solution : If the rotary switch is not set to “0”, the device will not function properly. Immediately turn
OFF the power and cease the operation. Contact our service representative.
Ventilator
The “Vent. Alarm” message is displayed.
Cause
: The following alarm has generated on the ventilator.
・
Parameter alarm such as AWP, MV, FiO
2
・
Technical alarm of the ventilator such as oxygen cell replacement
Solution : Check the alarm cause of the ventilator, and take appropriate action.
The “Vent. Data Unavailable” or “Vent. Offline” message is displayed. And the ventilator screen
is also displayed.
Cause 1 : The cable connecting the ventilator and DS-7000 is disconnected, or not securely
connected.
Solution : Check and make sure the cable between the ventilator and DS-7000 is connected and
secured.
Cause 2 : The power of the ventilator is turned OFF.
Solution : Turn the power of the ventilator ON.
Cause 3 : The ventilator is in standby mode.
Solution : Check if the ventilator is in ventilation mode, and take appropriate actions.
When normal communication is recovered, the communication status will be displayed as
“Vent.Online”.
10
-
26
Summary of Contents for DynaScope 7000 Series
Page 38: ...xxxvi Blank Page ...
Page 126: ...4 78 Blank Page ...
Page 148: ...5 22 Blank Page ...
Page 259: ...7 15 The Description of the Display Scale Real Time Scale Scale 7 Table Trend ...
Page 322: ...7 78 Blank Page ...
Page 428: ...10 32 Blank Page ...
Page 466: ...11 38 Blank Page ...
Page 474: ...12 8 Blank Page ...