Press the APNEA key to light the key indicator, and set the upper limit of the apnea alarm.
The set alarm limit will be applied to the apnea time
measured from the impedance respiration waveform,
ventilator, or multigas unit.
The upper limit can be set in 5-second increment. There
is no lower limit.
Key
Item
Description
Individual Setup
Selecting ON will generate the apnea alarm.
Selecting OFF will not generate the apnea alarm.
▲
Current Value
Indicates the current value. (–6 sec. for neonate.)
Upper Alarm
Limit
Sets the upper alarm limit (10 to 60sec.).
Setting the value equal to or above 60sec. will turn OFF
the alarm.
Automatic Setup
Sets the alarm limit to 60sec.
Also, “APNEA Step Alarm” will be set to ON.
Suspend Alarm
Suspends the RR and apnea alarm even if the
thresholds are exceeded.
5
Select ON/OFF for “APNEA Step Alarm”.
The “APNEA Step Alarm” selection will be displayed when the upper limit of apnea alarm is set to
30 seconds or above.
Selecting ON will generate level 3 alarm sound (single
beep tone) at 1/3 and 2/3 of the apnea upper alarm limit
along with an “Apnea Alarm” message.
To maintain the alarm setting even after the power is turned OFF or after the discharge procedure,
store the setting to one of the alarm modes, or select ON for “Backup at Discharge” on the
Discharge Mode setup menu under the Initial Settings.
Reference
Refer to “9. Initial Settings –Alarm Mode Setup–” for procedure to program the alarm mode.
6
Re
spi
ration
Summary of Contents for DynaScope 7000 Series
Page 38: ...xxxvi Blank Page ...
Page 126: ...4 78 Blank Page ...
Page 148: ...5 22 Blank Page ...
Page 259: ...7 15 The Description of the Display Scale Real Time Scale Scale 7 Table Trend ...
Page 322: ...7 78 Blank Page ...
Page 428: ...10 32 Blank Page ...
Page 466: ...11 38 Blank Page ...
Page 474: ...12 8 Blank Page ...