6
-
50
ON/OFF of Parameter Display
1
Press the Display ON/OFF key.
2
Select Display ON or Display OFF .
Pressing the Display ON key will display the waveform
and numeric data.
Pressing the Display OFF key will not display the
waveform and numeric data.
When SpO
2
sensor is attached to the patient with the SpO
2
display set to OFF, and SpO
2
is measured
for 10 seconds, the pulse wave and numeric data will be automatically displayed.
C A U T I O N
z
When the waveform and numeric data display is set to OFF, the alarm
generation and table/trend input will also be suspended.
z
When the waveform and numeric data display is set to OFF, the pulse
rate measured by SpO
2
will not be displayed either.
The “Disp. OFF” message will be
displayed inside the numeric data box.