C A U T I O N
y
When the waveform and numeric data display for the gas module is set
to OFF, the respiration rate measured by the gas module will not be
displayed.
y
If “Save Data to Table” is set to OFF on the alarm setup menu, data at
alarm occurrence will not be stored in the table.
y
Depending on the data stored in the table (alarm data, NIBP manual
measurement data, or NIBP continuous measurement data), the total
duration of the table will differ.
y
Do not set the same remote control bed ID to more than one monitors on
the same floor. Otherwise, it may cause to control more than one
monitors at the same time.
y
After the remote control setup, check that the remote control unit is
properly operating.
z
Patient Admit / Discharge
y
If you start monitoring a new patient without performing a discharge
procedure for the previous patient, new data will be added to the
previous data which will result in inaccuracy.
y
The setup for the alarm mode and display mode remains stored even
when the power is turned off or when discharging procedure is
performed. Before monitoring, make sure the current monitoring mode is
suitable for the patient’s condition.
y
On a wired network (DS-LANII/III) system, if the discharge procedure is
performed on the central monitor, the alarm setting on the bedside
monitor will be initialized to the value set on the “Admit Setup” of the
central monitor.
z
ST Measurement
y
For the lead which the electrode is detached, the reference waveform
setup cannot be performed. Check if the electrode is correctly attached,
and perform the setup again.
y
When the electrodes are properly attached, the reference waveform
setup will be automatically performed. To display the correct ST data, the
reference point and measurement point must be set.
z
CF Card
y
Use only the specified CF card.
y
Use only the CF card formatted with this device.
y
Restart the system after reading the setup data from the CF card.
The setup data will become effective after the system is restarted.
y
Reading the patient data from the CF card will erase all previous patient
data stored in the patient monitor.
z
Maintenance
y
If stains cannot be removed from the touch panel surface, wipe softly
with a dry or ethanol dampened cleaning cloth. Never use strong-acidic
cleaning solution. Neither is it recommended to use mild acidic or
alkaline cleaning solution.
y
A special coating is applied to the surface of the touch panel. Do not
wipe the surface with a cloth or gauze with coarse texture. Wipe the
surface with a soft cleaning cloth provided as optional accessory or with
an eyeglass cleaning cloth.
y
Clean the equipment frequently so stains can be removed easily.
y
To prevent injury, it is recommended to wear gloves when cleaning the
equipment.
y
Do not use organic solvents, thinner, toluene and benzene to avoid
damaging the resin case.
y
Do not polish the housing with abrasive or chemical cleaner.
y
When sterilizing the entire room using a spray solution, pay close
attention not to have liquids get into the equipment or connectors.
y
Use only neutral detergent to clean the housing. Do not use chemical
cloth, scrub brush, abrasive, polishing powder, hot water, volatile solvent
and chemicals (cleanser, thinner, toluene, benzine, benzol, and synthetic
detergent for house and furniture), or sharp-edged tools. The surface
resin coating may be damaged, resulting in discoloration, scratches, and
other problems.
xxiv
Summary of Contents for DynaScope 7000 Series
Page 38: ...xxxvi Blank Page ...
Page 126: ...4 78 Blank Page ...
Page 148: ...5 22 Blank Page ...
Page 259: ...7 15 The Description of the Display Scale Real Time Scale Scale 7 Table Trend ...
Page 322: ...7 78 Blank Page ...
Page 428: ...10 32 Blank Page ...
Page 466: ...11 38 Blank Page ...
Page 474: ...12 8 Blank Page ...