6
-
5
A message, “Zero failed” will be displayed if the process
fails. The three-way valve may not be opened to air, noise
interferenece may be present, or the transducer may be
defective.
Check the cause and try the zero balance process again.
A message, “Disconnect” will be displayed when the
interface cable is not connected. Check if the cable is
connected correctly.
4
When the zero balance process is complete, close the three-way valve.
C A U T I O N
Each time the blood pressure transducer or tubing is replaced, the zero
balance procedure is required to ensure accurate measurements
.
6
Zero Bal
an
ce
of Blood Pre
ssure