4
-
53
2
Basic Alarm will allow to set the alarm for basic parameters.
Other Alarm will allow to set the alarm for other parameters.
<
Basic Alarm Setup Menu
>
<
Other Alarm Setup Menu
>
3
Select ON/OFF and set upper and lower alarm limit for the parameter alarm.
【
Basic Alarm Setup
】
Numeric Data Key
Item
Description
HR/PR/BPR
ON, OFF 20 to 300bpm
NIBP
S
ON, OFF 60 to 250mmHg (Adult)
ON, OFF 40 to 180mmHg (Child)
ON, OFF 30 to 120mmHg (Neonate)
M ON, OFF 45 to 235mmHg (Adult)
ON, OFF 30 to 160mmHg (Child)
ON, OFF 15 to 100mmHg (Neonate)
D
ON, OFF 40 to 200mmHg (Adult)
ON, OFF 20 to 150mmHg (Child)
ON, OFF 10 to 90mmHg (Neonate)
SpO
2
ON, OFF 70 to 100%
TEMP1
ON, OFF 30 to 50
°
C
EtCO
2
ON, OFF 1 to 115mmHg
0.1 to 15.0kPa
0.1 to 15.0%
InspCO
2
(Upper Limit)
ON, OFF 1 to 24mmHg
0.1 to 3.0kPa
0.1 to 3.0%
RR
ON, OFF 5 to 150Bpm (Adult)
2 to 150Bpm (Child, Neonate)
APNEA
(Upper Limit)
ON, OFF 10 to 60sec.
4
Alarm Setup