Patient Sex
Select the patient’s sex from male or female. The default is set as undetermined. The selected sex will
be printed on the recording paper.
1
Select Male or Female .
This selection will not affect the measurement accuracy of the monitoring.
Pacemaker Usage
If the patient is wearing a pacemaker, the monitor will identify the pacemaker pulse and insert an
artificial pulse onto the ECG waveform for easy identification. By detecting the pacemaker pulse, it
prevents to erroneously detect QRS as a pacemaker pulse when pacing waveform does not appear
(pacing failure). The arrhythmia analysis analyzes pacing beat as P (pacemaker beat) or F (fusion
beat) to prevent erroneous judgment of VPC.
W A R N I N G
The pacemaker usage selection influences the precision of the QRS detection
and arrhythmia analysis. Make sure the correct selection is made.
1
Select Yes or No for pacemaker usage.
2
The pacemaker use will be displayed on the home display.
When pacemaker is used.
5
-
6