When the BP Label is IAP
When the BP label is IAP, PDP (Peak Diastolic Pressure) will be displayed in addition to systolic (S),
diastolic (D), and mean (M) pressure.
Please be aware that the Systolic Pressure (SYS) = Peak Systolic Pressure (PSP).
C A U T I O N
z
Note that the Systolic Pressure (SYS)
=
Peak Systolic Pressure (PSP)
for trend, data base, and alarm setup.
z
When ECG is not measured, PDP cannot be calculated.
●
When the BP Label is CVP
When the BP label is CVP, the measurement unit can be selected from “mmHg” or “cmH
2
O”.
The measurement unit can be selected on the preset menu. The selected unit will be displayed on the
BP numeric data box.
Reference
Refer to “9. Initial Settings –Measurement Unit–” for CVP measurement unit setup.
●
When the BP Label is ICP
When the BP label is ICP, labeling the artery pressure as ART will allow measuring the CPP (Cerebral
Perfusion Pressure). CPP = Mean Arterial Pressure
-
Intracranial Pressure.
If CPP is a negative value, the value will not be displayed. Also, the alarm cannot be set for CPP.