Cause 2 : “RR Alarm Source” is set to Ventilator .
Solution : Set the “RR Alarm Source” to Impedance .
N O T E
z
The impedance respiration waveform will not be displayed if GAS
is set for the
“RR Alarm Source”. AWF, AWP waveform will be displayed.
z
The CO
2
respiration waveform will not be displayed if Impedance
is set for
“RR Alarm Source”. AWF, AWP waveform will be displayed.
z
Both the CO
2
and impedance respiration waveform will not be displayed if
Ventilator
is set for “RR Alarm Source”.
Although the PR alarm is set on the central monitor, it returns to the previous setting after a
short time.
Cause 1 : HR is set for “HR/PR”.
Cause 2 : PR is set for “HR/PR”, but BP is set for “PR Source”.
Solution : Set PR for “HR/PR”, and set SpO
2
for “PR Source”
Although the HR alarm is set on the central monitor, it returns to the previous setting after a
short time.
Cause
: PR is set for “HR/PR”.
Solution : Set HR for “HR/PR”.
C A U T I O N
On the DS-7000, HR and PR alarm cannot be set to ON simultaneously.
Search Patient ID key does not function.
Cause 1 : The DS-LAN setup is set to “DS-LAN
Ⅱ
”.
Solution : Set
“DS-LAN
ΙΙΙ
” for all bedside monitors and central monitors in the same network.
When the setting is changed, restart the system.
Cause 2 : The communication with the central monitor is not established.
Solution : Check the communication with the central monitor.
Cause 3 : The communication between the central monitor and the patient data server is not
established.
Solution : Check the communication between the central monitor and the patient data server.
Cause 4 : The connected central monitor do not have the patient ID search function.
Solution : Check if the model type and software version of the central monitor supports the patient
ID search function.
The patient ID search process fails.
Cause 1 : The searched patient is not registered.
Solution : Check if the entered patient ID is correct.
Cause 2 : More than one bedside monitors are searching the same patient at the same time.
Solution : Try the search process again after a while.
10
-
24
Summary of Contents for DynaScope 7000 Series
Page 38: ...xxxvi Blank Page ...
Page 126: ...4 78 Blank Page ...
Page 148: ...5 22 Blank Page ...
Page 259: ...7 15 The Description of the Display Scale Real Time Scale Scale 7 Table Trend ...
Page 322: ...7 78 Blank Page ...
Page 428: ...10 32 Blank Page ...
Page 466: ...11 38 Blank Page ...
Page 474: ...12 8 Blank Page ...