Recommended Separation Distances between Portable and Mobile RF
Communications Equipment and the DS-7000
The DS-7000 is intended for use in an environment in which radiated RF disturbances are controlled.
The electromagnetic interference can be prevented by maintaining a minimum distance between
portable and mobile RF communications equipment (transmitters) and the DS-7000 as recommended
below, according to the maximum output power of the communications equipment.
Separation Distance according to Frequency of
Transmitter (m)
Rated Maximum
Output Power of
Transmitter
(
W
)
150kHz to 80MHz
d = 1.2
P
80MHz to 800MHz
d = 1.2
P
800MHz to 2.5GHz
d = 2.3
P
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
1 1.2
1.2
2.3
10 3.8 3.8 7.3
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in meters (m) can be determined using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
Note 1 : At 80MHz and 800MHz, the separation distance for the higher frequency range applies.
Note 2 : These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
xxxv
Summary of Contents for DynaScope 7000 Series
Page 38: ...xxxvi Blank Page ...
Page 126: ...4 78 Blank Page ...
Page 148: ...5 22 Blank Page ...
Page 259: ...7 15 The Description of the Display Scale Real Time Scale Scale 7 Table Trend ...
Page 322: ...7 78 Blank Page ...
Page 428: ...10 32 Blank Page ...
Page 466: ...11 38 Blank Page ...
Page 474: ...12 8 Blank Page ...