Precautions about the Ventilator Monitoring
W A R N I N G
z
The ventilator alarm on this monitor should be used as supplementary
function. Check the patient’s condition, ventilator alarm sound and
message occasionally.
z
The ventilator alarm sound is set to OFF at factory default setting.
The alarm sound can be turned ON on the volume setup menu.
z
If the DS-7000 does not generate an alarm even though the ventilator is
generating an alarm, or if any other malfunction occurs, immediately
check the ventilator, DS-7000, cable, and replace the cable if necessary.
If the malfunction persists, stop using the device.
z
After connecting the ventilator and the DS-7000, ensure that “Vent.
Online” message is displayed for the connection status. Otherwise, the
DS-7000 will not detect the ventilator alarm.
z
The alarm generation on the DS-7000 is not assured if the alarm other
than specified generates on the Servo-i.
y
airway pressure upper limit alarm, high continuous pressure alarm, O
2
concentration lower limit alarm, expiratory minute volume upper/lower
limit alarm, apnea alarm, gas supply alarm, O
2
supply alarm, battery
alarm, no battery alarm, limited battery alarm, overrange alarm,
expiratory cassette disconnected alarm, backup ventilation alarm,
regulation pressure limited alarm, respiratory rate alarm, PEEP low
alarm, EtCO
2
upper limit alarm, EtCO
2
lower limit alarm
C A U T I O N
z
The ventilator operation should be performed by well-trained and
authorized personnel.
z
For connecting the DS-7000 and ventilator, use only the specified
connection cable.
z
Verify that the DS-7000 and the ventilator are properly connected.
z
When connecting the cable, verify that the main power of the DS-7000
and the ventilator is OFF.
z
Check occasionally the communication status of the DS-7000 and the
ventilator.
z
Verify that the ventilator alarm is not generated, and the “Vent. Online”
message is displayed.
z
The “Check external alarm” will be displayed until proper communication
with the ventilator is resumed. When the communication is resumed, the
screen will automatically return to the home display.
z
When disconnecting the ventilator and the DS-7000, make sure to select
OFF on the “Check external alarm” display which appears when the
power of the ventilator is turned OFF, or when the cable is disconnected.
Precautions for Use of SpO
2
Sensor
D A N G E R
Burn Risk in Using SpO
2
Sensor
In SpO
2
monitoring, always use the sensor/relay cable specified by
Fukuda Denshi. If any other sensor/relay cable is used, a high temperature
rise of the sensor may place the patient in danger of burns.
If there are any questions regarding the sensor/relay cable use for SpO
2
measurements of this device, please contact Fukuda Denshi service
representative.
xxx
Summary of Contents for DynaScope 7000 Series
Page 38: ...xxxvi Blank Page ...
Page 126: ...4 78 Blank Page ...
Page 148: ...5 22 Blank Page ...
Page 259: ...7 15 The Description of the Display Scale Real Time Scale Scale 7 Table Trend ...
Page 322: ...7 78 Blank Page ...
Page 428: ...10 32 Blank Page ...
Page 466: ...11 38 Blank Page ...
Page 474: ...12 8 Blank Page ...