To Perform Arrhythmia Learning
Learning the normal ECG largely affects the accuracy of the arrhythmia analysis.
If any error occurs in arrhythmia detection and QRS judgment, performing arrhythmia learning will
recover the original analyzing accuracy. Arrhythmia learning will be performed for about 20 beats for a
normal ECG, but it may take longer if the heartbeat is unstable.
During arrhythmia learning, arrhythmia alarm other than ASYSTOLE, TACHY, BRADY will not be
generated.
1
Press the Menu
→
Settings
→
Parameter
→
ECG keys.
2
Start arrhythmia learning.
Arrhythmia learning is in progress.
Pressing the key with the LED off will start arrhythmia learning.
Arrhythmia learning process cannot be stopped by pressing the key again.
3
During arrhythmia learning, the message “LEARN” will be displayed.
N O T E
If a pacemaker is used, the LED on the Arrhythmia Learn key will not light and the
“LEARN” message will not be displayed although the arrhythmia learn procedure is
performed.
Pressing the HR parameter key will also display
the ECG parameter setup menu.