End of Measurement Tone
By selecting ON for the “End Tone”, a tone will be generated when the NIBP measurement completes.
There are 3 types of “End Tone” depending on the measurement result comparison with the previous
value.
If SYS > Previous SYS,
When previous SYS >= 101 and (SYS – previous SYS) >= 15mmHg, or
When previous SYS <= 100 and (SYS – previous SYS) >= 10mmHg,
the “End Tone” will generate with an increasing tone (high tone) to notify an increase in the BP.
If SYS < Previous SYS,
When previous SYS >= 101 and (previous SYS – SYS) >= 15mmHg, or
When previous SYS <= 100 and (previous SYS – SYS) >= 10mmHg,
the “End Tone” will generate with a decreasing tone (low tone) to notify a decrease in the BP.
For condition other than the above, a medium tone will be generated.
1
Select ON or OFF for “End Tone”.
ON will generate a tone when the measurement
completes.
OFF will not generate a tone when the measurement
completes.
●
Pump Setup
The pump inflation operation can be selected from normal or silent operation.
1
Display the second page of the NIBP setup menu to set the “Pump Setup”.
Silent will inflate with decreased speed to reduce the
pump inflating sound.