GE
Measurement & Control
Druck DPI611
Portable Pressure Calibrator
Safety and Quick Start Guide
–
K0565
DPI611: Pressure Calibrator (Figure 1)
Maximum Operating Pressure:
As stated on Range
Label at the back side of Calibrator.
Over pressure limited to 1.5 × MWP
Note: MWP is based on Unit pressure range
Operating Temperature Range
:
-10 °C to +50 °C (+14 °F to +122
°
F)
+10 °C to +30 °C (optimized calibration specification)
0 °C
–
40 °C with optional mains power supply
Storage Temperature:
-20 °C to +70 °C (-4 °F to +158 °F)
Humidity
: 0 to 90 % Relative Humidity (RH)
Non-condensing
Power Supply:
100
–
260 V +/-10 %, 50 / 60 Hz AC,
Output DC V=5 A, 1.6 A, CATII of IEC60364-4-443
Pressure Ranges:
+/-1 barg (-14.5 to 15 psig)
-1 to 2 barg (-14.5 to 30 psig)
-1 to 7 barg (-14.5 to 100 psig)
-1 to 10 barg (-14.5 to 150 psig)
-1 to 20 barg (-14.5 to 300 psig)
Accuracy 10 °C to 30 °C, 1 yr. 0.025 % FS
Add 0.001 % FS/°C outside 10 °C to 30 °C
Altitude:
Up to 2000 m
Pollution Level:
Degree 2
Batteries (Primary):
8 × AA Alkaline
*Caution: To prevent damage to the DPI611, only use
it within the specified pressure limit on the label.
© 2014 General Electric Company. All rights reserved.
Trademarks
All product names are trademarks of their respective companies.