background image

 

 

 

Measurement & Control

 

 

 

[EN] English 

K0565 Issue 1  

Safety and Quick start guide 

3

 

 

ELECTRICAL WARNINGS 

 

 

 

 

To  prevent  electrical  shocks  or  damage  to  the  instrument,  do  not 
connect  more  than  30 V  between  the  terminals,  or  between  the 
terminals and the ground (earth). 
 

 

This  instrument  uses  standard  AA  size  batteries.  To  prevent  an 
explosion or fire, do not short circuit.

 

 

 

To  prevent  battery  leakage  or  heat  generation,  only  use  the  power 
supply in the temperature range 0 to 40 °C (32 to 104 °F). 

 
PRESSURE WARNINGS 
 

 

 

 

Some  liquid  and  gas  mixtures  are  dangerous.  This  includes  mixtures 
that occur because of contamination. Make sure that the equipment is 
safe to use with the necessary media. 
 

 

To prevent a dangerous release of pressure, isolate and bleed the 
system before you disconnect a pressure connection. 

 

 

To  prevent  a  dangerous  release  of  pressure,  make  sure  that  all  the 
related pipes, hoses and  equipment have the correct pressure rating, 
are safe to use and are correctly attached. 
 

 

Always wear appropriate eye protection when working with pressure. 

 

 

 

Summary of Contents for Druck DPI611

Page 1: ...o 158 F Humidity 0 to 90 Relative Humidity RH Non condensing Power Supply 100 260 V 10 50 60 Hz AC Output DC V 5 A 1 6 A CATII of IEC60364 4 443 Pressure Ranges 1 barg 14 5 to 15 psig 1 to 2 barg 14 5...

Page 2: ...FRENCH FR GERMAN DE ITALIAN IT SPANISH ES PORTUGUESE PT RUSSIAN RU CHINESE ZH GREEK EL CZECH CS DANISH DA HUNGARIAN HU LITHUANIAN LT LATVIAN LV SWEDISH SV DUTCH NL NORWEGIAN NO POLISH PL TURKISH TR R...

Page 3: ...for Voltage V 24 V loop power supply 24 Vo 8 Liquid Crystal Display LCD Color display with touch screen To make a selection lightly tap on the applicable display area 9 5 V DC power input socket This...

Page 4: ...7 bar 5 mbar min QUICK REFERENCE Before using this instrument read and understand the Safety section It is dangerous to ignore the specified warnings This instrument is intended for indoor use only GE...

Page 5: ...upply in the temperature range 0 to 40 C 32 to 104 F PRESSURE WARNINGS Some liquid and gas mixtures are dangerous This includes mixtures that occur because of contamination Make sure that the equipmen...

Page 6: ...t parameters The DPI611 instrument measures and displays Pneumatic Pressure Vacuum generated by the instrument In addition it can measure source electrical signals 2 STANDARD EQUIPMENT These items are...

Page 7: ...18 4 lbf ft Attach the pressure adaptor to the DPI611 and tighten to finger tight 4 3 Maintenance Clean the case with a moist lint free cloth and a weak detergent Do not use solvents or abrasive mate...

Page 8: ...ing the cover upwards 2 Remove old batteries and dispose off in accordance with local instructions 3 Place the new batteries in the battery compartment with the correct orientation 4 Replace the batte...

Page 9: ...ssing and holding the power button until the screen is blank 2 Use Power OFF to preserve the battery capacity when the instrument is not in use for extended periods of time Standby Mode Press and Rele...

Page 10: ...e DASHBOARD displays icons representing the functional applications of the DPI611 for example the CALIBRATOR application Touch the appropriate icon to start the application When more choices are avail...

Page 11: ...USTOM TASK and manually configure the required measure and source functions CUSTOM TASK tasks can be saved for future use by adding them to FAVOURITES using the Save function in the TASK MENU 7 1 Set...

Page 12: ...e the applicable adaptor to attach the device See Figure 9 Step Procedure Vacuum 1 Set to vacuum operation 2 Turn the volume adjuster to mid range 3 Seal the system 4 Set the vacuum with the pump 5 Ad...

Page 13: ...C BS EN 61326 1 2013 Electrical safety BS EN 61010 1 2010 Pressure safety Pressure Equipment Directive Class Sound Engineering Practice SEP Approved CE Marked Complies with European Union directives W...

Page 14: ...Safety and Quick start guide 12 EN English K0565 Issue 1...