![background image](http://html1.mh-extra.com/html/ge/druck-dpi611/druck-dpi611_quick-start-manuals_4133942007.webp)
Measurement & Control
[EN] English
–
K0565 Issue 1
Safety and Quick start guide
5
4.
I
NSTALLATION
:
Before you start:
Read and understand the “Safety” section
.
Do not use damaged equipment.
Note:
Use only original parts supplied by the manufacturer.
4.1
Electrical Connections
See Figure 1 (View A).
4.2
External pressure connections
See Figure 2. Use an applicable method to seal the external
pressure connections, and then tighten to the applicable torque.
Maximum torque:
ISO 228/1, 1/8 NPT: 35 Nm (26 lbf.ft)
ISO 228/1, G1/8: 25 Nm (18.4 lbf.ft)
Attach the pressure adaptor to the DPI611 and tighten to finger
tight.
4.3
Maintenance
Clean the case with a moist, lint-free cloth and a weak detergent.
Do not use solvents or abrasive materials.
Return the instrument to the manufacturer or an approved service
agent for all repairs. Refer to the user manual.
If the equipment is used in a manner not specified by the
manufacturer, the protection provided by the equipment may be
impaired.
Do not dispose of this product as household waste. Use an
approved organization that collects and/or recycles waste
electrical and electronic equipment. For more information, contact:
Our customer service department:
(Contact us at
www.ge-mcs.com
).