5-32
Troubleshooting the Analyzer
Both main fuses, F1 and F2, must be replaced if either one is blown.
Location:
Two main fuses,
F1
and
F2
(
1
); fuses
F3
through
F7
(
2
)
Analyzer mode: Off, power cord unplugged
To replace the 20-mm fuses
(F3 through F7)
1.
Loosen the fuse holder by turning it
counterclockwise.
2.
Pull out the fuse holder, then remove
the faulty fuse.
3.
Insert the new fuse, slide the fuse
holder in, and tighten it by turning it
clockwise.
4.
Plug in the power cord and set the
power switch to
I
(on).
To replace main fuses
(F1 and F2)
NOTE
Both fuses must be replaced.
1.
With a screwdriver, loosen the fuse
cap by turning it counterclockwise.