Flags
6-21
Sample/System Flags
Summary
Through the use of complex flagging algorithms, laboratory personnel are alerted
to suspected abnormal sample and/or system conditions.
Sample/system flags appear on the Run Screen and the Review / Edit tab.
Whenever such flags are triggered, the user should review the results and take the
action recommended. The following table identifies the sample/system flags that
could be triggered for each test selectivity. The two-letter codes in parentheses
appear on the Sample/System Flag Analysis panel of the Run Screen.
Flag
Triggering Criteria
CBC
CBC /
DIFF
CBC / DIFF
/ RETIC
CBC /
RETIC
RETIC
B-SUSP (BC) Events above the upper
MN/PMN threshold on y
axis and in channels 0 to 49
on x axis
≥
5%
BIFR (BR)
Baso Flatness > 2.5
B-NO (NB)
Baso % Noise > 10
B-NV (VB)
1.
(BASO d/D) < 0.15
2.
Flag is not set if there is
WBC agreement (no
WBC-CE flag) and
[(%NEUT + %EOS) -
%PMN] is between 0
and 7.5.
B-SAT (BS)
1.
Baso % Saturation > 2.5
2.
Flag is not set if there is
WBC agreement (no
WBC-CE flag).
BTO (TB)
1.
Baso chamber
temperature is not
between 31.9°C and
34.1°C.
2.
Flag is not set if there is
WBC agreement (no
WBC-CE flag) and
[(%NEUT + %EOS) -
%PMN] is between 0
and 7.5.
CHCMCE
(CC)
(MCHC - CHCM) > 1.9
Summary of Contents for ADVIA 2120
Page 1: ...ADVIA 2120 2120i Hematology Systems Operator s Guide 067D0157 01 Rev 201 0 ...
Page 30: ...1 26 Welcome to the ADVIA 2120 2120i Hematology System ...
Page 46: ...3 12 Daily Routine ...
Page 171: ...7 2 Status Line messages T 93 U 94 V 96 W 98 X 99 Y 99 Z 99 ...
Page 269: ...7 100 Status Line messages ...
Page 341: ...8 72 Methods ...