1-12
Welcome to the ADVIA 2120/2120i Hematology System
Reagents, SHEATH / RINSE, Wash, and DEFOAMER
Reagent
is mixed with a sample for cytochemical analysis and measurement.
Sheath
is a fluid that envelops the sample stream as it passes through the optics,
ensuring cell-by-cell analysis.
Rinse
cleans the hydraulic pathways and reaction chambers after each sample to
prevent carryover and ensure the integrity of the results.
Wash
solution removes buildup in the hydraulic pathways. A system wash
should be performed periodically: once a day or after a set number of samples.
DEFOAMER
reduces foam buildup in the waste container.
Wash, PEROX SHEATH, DEFOAMER, and all cytochemical reagents are
located on the analyzer and, except for the SHEATH / RINSE and DEFOAMER,
they are monitored on a cycle-by-cycle basis.
The SHEATH / RINSE is in a CUBITAINER
®
placed on the floor.
Vacuum and Pressure Regulator Knobs
20 In.-Hg
Vacuum Regulator
Knob
40 PSI
Pressure Regulator
Knob
5 PSI
Pressure Regulator
Knob
20 PSI
Pressure Regulator
Knob
Touchpad
Start/Stop
Sampler
Eject
Rack
Rack
in Sampler
Standby
On
Off
The touchpad, located on the lower right of the analyzer, is the means by which
you operate the analyzer.
On –
Provides power to the analyzer assemblies and power supplies. After
On
is
pressed, samples through the manual open- and closed-tube samplers can be
processed within 2.5 minutes, while samples through the autosampler can begin
processing within four minutes.
Standby –
Sets the analyzer to a lower power state. To exit, press
Standby
.
Summary of Contents for ADVIA 2120
Page 1: ...ADVIA 2120 2120i Hematology Systems Operator s Guide 067D0157 01 Rev 201 0 ...
Page 30: ...1 26 Welcome to the ADVIA 2120 2120i Hematology System ...
Page 46: ...3 12 Daily Routine ...
Page 171: ...7 2 Status Line messages T 93 U 94 V 96 W 98 X 99 Y 99 Z 99 ...
Page 269: ...7 100 Status Line messages ...
Page 341: ...8 72 Methods ...