Troubleshooting the Analyzer
5-9
•
Household bleach
Time: 4 minutes
Analyzer mode: Ready to Run
1.
Open the front cover.
2.
At the Utilities menu, select the Exerciser tab, and select the
Valves
button
on the left.
3.
Remove the tubing from the
V45
nipple on the autosampler centering collar.
V43
V44
V45
4.
Place the end of the tube into a beaker of 25% solution of household bleach
and water.
5.
At the Exerciser
,
select
V45
to open the valve.
6.
Aspirate 10 mL of the bleach, followed by 10 mL of deionized water
7.
At the Exerciser, select
V45
to close the valve.
8.
Replace the tubing on the centering collar nipple, and exit the
Exerciser
window.
Cleaning the Vent Lines, Vacuum Shuttle Chambers, and
Reaction Chambers
Materials required:
•
Beaker
•
Deionized water
•
EZ KLEEN for cleaning perox-related pathways and chambers
•
Household bleach for all pathways and chambers other than perox
•
Tubing, 0.090, PN 116-0536-16
Time: 4 minutes
Analyzer mode: Ready to Run