7-76
Status Line messages
Possible Blasts or Abnormal Cells - Stop
The number of consecutive occurrences of this condition has met the criterion
specified for a stop message. The autosampler halts at the same time the stop
message occurs. The specified stop criterion is user definable in the Customize
Menu, System Setup, Alarm/Stop Criteria. The counter associated with the Stop
Criteria is reset when the Start/Stop button is selected.
Power Pack Temperature High
The internal temperature of the power supply has exceeded the maximum limit.
Possible Cause
1
Fan filters are
dirty
Corrective Action
Shut down the system and remove and clean the air-
circulation filter as directed in the Operator's Guide.
2
System internal
temperature
exceeds the
maximum limit
a
Shut down the system.
b
Allow the system to cool down for 15 minutes.
c
Restart the system.
d
If the problem persists, call Siemens Service for
assistance.
3
Fans are not
functioning
properly
Verify fan operation. If you suspect that a fan is not
working properly, call Siemens Service for assistance.
4
Faulty power
pack
temperature
sensor
Call Siemens Service for assistance.
Pressure Node Error - Reset Required
Potential software communication error or a hardware failure associated with the
Pressure Node Board.
Possible Cause
1
Hardware
failure
Corrective Action
a
Reset the analyzer by selecting
Utilities, Exerciser,
Indicators, Analyzer Reset
.
b
After about 1 minute, if the status line does not
display Ready to Run and the green ready-to-aspirate
light is still off, follow these steps to exit the ADVIA
2120/2120i software, turn off the analyzer and then
restart.
Summary of Contents for ADVIA 2120
Page 1: ...ADVIA 2120 2120i Hematology Systems Operator s Guide 067D0157 01 Rev 201 0 ...
Page 30: ...1 26 Welcome to the ADVIA 2120 2120i Hematology System ...
Page 46: ...3 12 Daily Routine ...
Page 171: ...7 2 Status Line messages T 93 U 94 V 96 W 98 X 99 Y 99 Z 99 ...
Page 269: ...7 100 Status Line messages ...
Page 341: ...8 72 Methods ...