8-36
Methods
FracDT
The fraction of time that the channel is busy processing flowcell events. While
the perox channel is busy identifying a particular flowcell event, it is unable to
process any additional events that might occur. By measuring this "dead time",
the analyzer can compensate for these events.
MPXI
The mean peroxidase activity index or staining intensity of the neutrophil
population relative to the archetype.
PEROX PHA Cells (P-acq)
The total number of events in the PEROX cytogram excluding the Noise area.
NOTE:
nRBC and PLT Clump events are components of Perox Noise. Therefore,
they are not included in the PEROX PHA Cells.
PEROX PHA Total (P-tot)
The total number of events in the PEROX cytogram including the Noise area.
Perox Valid Cells (P-vc)
The number of valid electronic pulse signals detected from flowcell events.
HPX
The number of events with absorption values greater than 1.4 times the X
channel mean of the neutrophil cluster.
Clumps Count
The number of events in the
Platelet Clumps area (1) of
the PEROX cytogram.
Valley Count
The number of events in the
NRBC area (1) of the
PEROX cytogram.
Summary of Contents for ADVIA 2120
Page 1: ...ADVIA 2120 2120i Hematology Systems Operator s Guide 067D0157 01 Rev 201 0 ...
Page 30: ...1 26 Welcome to the ADVIA 2120 2120i Hematology System ...
Page 46: ...3 12 Daily Routine ...
Page 171: ...7 2 Status Line messages T 93 U 94 V 96 W 98 X 99 Y 99 Z 99 ...
Page 269: ...7 100 Status Line messages ...
Page 341: ...8 72 Methods ...