Troubleshooting the Analyzer
5-39
2.
Unscrew the securing cap to which the
power cable (
2
) is attached. Carefully pull
the lamp assembly out of the optics lamp
housing. The lamp assembly includes the
lamp, the power cable, and the securing
cap. Discard the entire assembly.
VIEW FROM
BACK OF PEROX
OPTICS ASSEMBLY
3.
Insert the replacement lamp assembly.
CAUTION
Do not touch the glass bulb. If necessary, clean with lens paper and methyl
alcohol.
4.
Replace the securing cap, then connect the red power connector.
5.
Turn the analyzer on and align the lamp to maximize its output.
(See page 5-4 for instructions.)
Replacing the perox flowcell
Time: Installation - 15 minutes
Checkout - 15 minutes
Analyzer mode: Standby
BIOHAZARD
All products or objects that come in contact with human or animal body fluids
should be handled, before and after cleaning, as if capable of transmitting
infectious diseases. Wear facial protection, gloves, and protective clothing.
The operator should follow the recommendations to prevent the transmission of
infectious agents in health-care settings as recommended for potentially
infectious specimens in
Protection of Laboratory Workers from Infectious
Disease Transmitted by Blood, Body Fluids, and Tissue
, 2nd edition; Approved
Guideline (1997) Document M29-A, National Committee for Clinical Laboratory
Standards (NCCLS). This document contains complete information on user
protection and it can be used as reference material for instructions on laboratory
safety.
IMPORTANT
An inspection scope must be available for this procedure. Do not remove the
flowcells unless you are trained in how to align and focus them.
To remove the flowcell
1.
Make sure that the analyzer is in the Standby mode.
Summary of Contents for ADVIA 2120
Page 1: ...ADVIA 2120 2120i Hematology Systems Operator s Guide 067D0157 01 Rev 201 0 ...
Page 30: ...1 26 Welcome to the ADVIA 2120 2120i Hematology System ...
Page 46: ...3 12 Daily Routine ...
Page 171: ...7 2 Status Line messages T 93 U 94 V 96 W 98 X 99 Y 99 Z 99 ...
Page 269: ...7 100 Status Line messages ...
Page 341: ...8 72 Methods ...