Status Line messages
7-63
No Type
The type of test request sent by the host is not defined in the Test Dictionary.
Corrective Action
Verify the accuracy of the host test table. Define the test in Test Dictionary.
Select Customize menu, System Setup, Tools Modify, Test Dictionary.
No Workorder - Alarm
No workorder received for current patient sample. The number of consecutive
occurrences of this condition has met the criterion specified for an alarm
message.
Possible Cause
1
A work order
has not been
created for the
current patient
sample.
Corrective Action
Verify that a work order has been created at the host
computer or at the Data Manager.
2
A work order
has not been
transmitted for
the current
patient sample.
Check if problem exists at the host computer. If problem
is not at the host level, call Siemens Service for
assistance.
No Workorder - Stop
No workorder received for current patient sample. The number of consecutive
occurrences of this condition has met the criterion specified for a stop message.
The autosampler has stopped.
Possible Cause
1
A work order
has not been
created for the
current patient
sample.
Corrective Action
Verify that a work order has been created at the host
computer or at the Data Manager.
2
A work order
has not been
transmitted for
the current
patient sample.
Check if problem exists at the host computer. If problem
is not at the host level, call Siemens Service for
assistance.
Summary of Contents for ADVIA 2120
Page 1: ...ADVIA 2120 2120i Hematology Systems Operator s Guide 067D0157 01 Rev 201 0 ...
Page 30: ...1 26 Welcome to the ADVIA 2120 2120i Hematology System ...
Page 46: ...3 12 Daily Routine ...
Page 171: ...7 2 Status Line messages T 93 U 94 V 96 W 98 X 99 Y 99 Z 99 ...
Page 269: ...7 100 Status Line messages ...
Page 341: ...8 72 Methods ...