5-48
Troubleshooting the Analyzer
4.
Mark the tubes with tape (the top line is rinse and the bottom line is vacuum),
then remove the lines.
Be careful not to allow the tubes to fall behind the
aspirate paddle assembly. If necessary, use hemostats to hold the tubing
outside the assembly.
5.
Connect the lines to the new block.
6.
Install the block onto the aspirator assembly by aligning the tabs with the
appropriate slots. Push the block into the assembly and gently slide the block
upward until it clicks into place. Be careful that you do not pinch the wash or
vacuum line during the installation process.
7.
Check the length of the probe. (See page 5-3 for instructions.)
8.
Run several saline samples and verify that there are no leaks or bubbles in
the sample line.
9.
Run a whole blood primer to verify that the probe is being properly washed
and dried after aspiration.
10.
Run controls to verify analyzer performance.