7-32
Status Line messages
d
Replace autosampler CPU board.
Autosampler - Output Queue Jammed
The output queue is jammed.
Possible Cause
1
Leading edge
of the output-
queue pan
blade spears
rack.
Corrective Action
Remove racks from output queue and/or output shuttle.
2
Output-queue
motion
impeded.
a
Check for foreign objects in output queue, queue
covers, or pinch guard.
b
Make sure output queue moves freely.
3
Feed-motor
failure.
Call Siemens Service for assistance.
Autosampler - Output Queue Motion Denied
To prevent the collision of autosampler components, the system did not execute
the command to move the autosampler output queue.
Autosampler - Output Queue Motion Failed
The autosampler output queue failed to move properly.
Possible Cause
1
Leading edge
of the output-
queue blade
has speared a
rack.
Corrective Action
a
Remove racks from output queue and shuttle.
b
Select
Start
at the analyzer touchpad. The autosampler
resumes operation.
2
Queue cannot
move
properly.
Check the output queue, queue covers, and pinch guard
for foreign objects. Make sure the queue moves freely.
3
Feed motor
has failed.
Call Siemens Service for assistance.
Autosampler - Output Shuttle Blocked
A rack in the output-shuttle position is preventing the autosampler from
completing a movement.
Summary of Contents for ADVIA 2120
Page 1: ...ADVIA 2120 2120i Hematology Systems Operator s Guide 067D0157 01 Rev 201 0 ...
Page 30: ...1 26 Welcome to the ADVIA 2120 2120i Hematology System ...
Page 46: ...3 12 Daily Routine ...
Page 171: ...7 2 Status Line messages T 93 U 94 V 96 W 98 X 99 Y 99 Z 99 ...
Page 269: ...7 100 Status Line messages ...
Page 341: ...8 72 Methods ...