Turning the System On / Off
2-3
in progress. When the system preparation is complete, log on to the
system.
During system preparation, the analyzer:
•
Performs internal diagnostics checks
•
Prepares the hydraulics
•
Primes the reagent lines
•
Begins the Startup process
NOTE
If any of the internal diagnostic checks fail, the system displays an error
message on the status line. Turn off the system, wait 60 seconds, and
then restart it. If the error recurs, call your Service Representative.
c.
The system automatically opens the Startup tab. Review the status of the
Startup process.
IMPORTANT
In cases when the analyzer is turned off
and restarted too quickly, the compressor
may not start up, causing the system
initialization to fail. This is due to
residual vacuum in the waste container.
The operator must manually vent the
container.
To manually vent either container (manual or autowaste), disconnect the waste
container vacuum line (
1
), wait for the vacuum to dissipate, and then reconnect
the line.