
Status Line messages
7-93
System Preparation in Progress
When the ADVIA 2120/2120i is first powered on or the software is restarted, the
analyzer and the computer go through a preparation process which includes the
following:
•
Software initialization
•
Internal diagnostic checks
•
Priming of reagent lines
•
Startup, including a background verification cycle
System Preparation Not Complete
Corrective Action
a
Reset the analyzer by selecting
Utilities, Exerciser, Indicators, Analyzer
Reset
.
b
After about 1 minute, if the status line does not display
Ready to Run
and
the green ready-to-aspirate light is still off, follow these steps to exit the
ADVIA 2120/2120i software, turn off the analyzer and then restart.
i
Select
Shut Down ADVIA
at the Log On/Off window.
ii
Select
Off
at the analyzer touchpad.
iii
Select
CTRL+ALT+DELETE
and log back onto the system to restart the
software.
iv
Restart the analyzer by selecting
On
at the analyzer touchpad.
T
Test Not Buffered - cTestCode
There are more than 60 tests ACTIVE in the Test Dictionary. No additional tests
can be stored in memory.
Corrective Action
Cancel the selection of the active test that exceeded this limit in the Test
Dictionary. Select Customize menu, System Setup, Tools Modify, Test
Dictionary.
Tube xxx Not Transmitted to the Host
An expected result was not transmitted.
Corrective Action
Summary of Contents for ADVIA 2120
Page 1: ...ADVIA 2120 2120i Hematology Systems Operator s Guide 067D0157 01 Rev 201 0 ...
Page 30: ...1 26 Welcome to the ADVIA 2120 2120i Hematology System ...
Page 46: ...3 12 Daily Routine ...
Page 171: ...7 2 Status Line messages T 93 U 94 V 96 W 98 X 99 Y 99 Z 99 ...
Page 269: ...7 100 Status Line messages ...
Page 341: ...8 72 Methods ...