6-16
Flags
Possible Causes
The following list of sample-related causes may not contain all conditions that
could cause this flag.
Multiple occurrences of this flag, especially for consecutive samples, can indicate
a system problem. However, isolated instances of this flag are usually sample
related.
Sample Related
System Related
•
Myeloperoxidase
deficiency
1.
Troubleshoot the perox channel.
2.
Check the perox reagents.
3.
Check peroxidase hydraulics.
4.
Check the perox reaction chamber
temperature.
Nucleated Red Blood Cells (NRBC)
Definition
The detection of nucleated red blood cells, or the occurrence of either the Suspect
Cellular Interference (NRCELL) flag or the Suspect Large Platelets (NRLPLT)
flag will trigger the NRBC Morphology flag.
There is no severity level associated with this flag.
Possible Causes
The following list of sample-related causes may not contain all conditions that
could cause this flag.
Multiple occurrences of this flag, especially for consecutive samples, can indicate
a system problem. However, isolated instances of this flag are usually sample
related.
Sample Related
System Related
•
Neonatal samples·
•
Hemoglobinopathies·
•
Hemolytic anemia·
•
Myelodysplastic
syndromes
1.
Check perox reagents.
2.
Check the perox reaction chamber
temperature.
3.
Check peroxidase hydraulics and sample
delivery.
4.
Check perox gains.
Summary of Contents for ADVIA 2120
Page 1: ...ADVIA 2120 2120i Hematology Systems Operator s Guide 067D0157 01 Rev 201 0 ...
Page 30: ...1 26 Welcome to the ADVIA 2120 2120i Hematology System ...
Page 46: ...3 12 Daily Routine ...
Page 171: ...7 2 Status Line messages T 93 U 94 V 96 W 98 X 99 Y 99 Z 99 ...
Page 269: ...7 100 Status Line messages ...
Page 341: ...8 72 Methods ...