7-6
Status Line messages
5
5 PSI Out of Range
System pressure is out of range.
Possible Cause
1
5 PSI pressure
gauge is out of
range
Corrective Action
Check the 5 PSI gauge and adjust the regulator if
necessary.
Setpoint: 5 PSI
±
0.5
2
Pressure leak
Check for crimped or disconnected hydraulic or
pneumatic lines. Straighten or reconnect lines as
necessary.
5 PSI Out of Range - Stop
System pressure is out of range. The system has stopped because the pressure has
reached a critical level that may affect the integrity of sample results and damage
the system.
Possible Cause
1
5 PSI pressure
gauge is out of
range
Corrective Action
Check the 5 PSI gauge and adjust the regulator if
necessary.
Setpoint: 5 PSI
±
0.5
2
Pressure leak
Check for crimped or disconnected hydraulic or
pneumatic lines. Straighten or reconnect lines as
necessary.
IMPORTANT
If problems persist, Call Siemens Service for assistance. Further corrective action
must be taken by Siemens service personnel only.
8
80 Conflict
A conflict has occurred on a sample that was accessed by two different tasks (for
example, Review and Edit and Host Communication), or you attempted to access
a sample at the same time the Data Manager was attempting to access the sample.
Corrective Action
Wait and access the sample again a few seconds later.
9
9 End of File
The Control Dictionary is empty. Quality Control cannot be operated.
Summary of Contents for ADVIA 2120
Page 1: ...ADVIA 2120 2120i Hematology Systems Operator s Guide 067D0157 01 Rev 201 0 ...
Page 30: ...1 26 Welcome to the ADVIA 2120 2120i Hematology System ...
Page 46: ...3 12 Daily Routine ...
Page 171: ...7 2 Status Line messages T 93 U 94 V 96 W 98 X 99 Y 99 Z 99 ...
Page 269: ...7 100 Status Line messages ...
Page 341: ...8 72 Methods ...