Status Line messages
7-87
Running
Displayed after pressing the autosampler Start/Stop button or during sample
processing.
Possible Cause
If the system halts
in the
Running
state, there could
be a hardware or
software failure.
Corrective Action
a
Reset the analyzer by selecting
Utilities, Exerciser,
Indicators, Analyzer Reset
.
b
After about 1 minute, if the status line does not
display Ready to Run and the green ready-to-aspirate
light is still off, follow these steps to exit the ADVIA
2120/2120i software, turn off the analyzer and then
restart.
i
Select
Shut Down ADVIA
at the Log On/Off
window.
ii
Select
Off
at the analyzer touchpad.
iii
Select
CTRL+ALT+DELETE
and log back onto
the system to restart the software.
iv
Restart the analyzer by selecting
On
at the
analyzer touchpad.
S
# $ > Sample ID Not Read - Alarm
Sample-tube barcode label is unreadable or missing. The number of consecutive
occurrences of this condition has met the criterion specified for an alarm
message.
Possible Cause
1
Barcode label
improperly
placed,
obscured, or
not present.
Corrective Action
Verify that a barcode is present and correctly applied to
the sample tube (with the label flush against the rim of
the stopper). Verify that the barcode is clean and of
specified quality.
2
Barcode reader
faulty.
Call Siemens Service for assistance.
Sample ID Not Read - Stop
Sample-tube barcode label is unreadable or missing. The number of consecutive
occurrences of this condition has met the criterion specified for a stop message,
and the autosampler has halted.
Possible Cause
1
Barcode label
improperly
Corrective Action
Verify that a barcode is present and correctly applied to
the sample tube (with the label flush against the rim of
Summary of Contents for ADVIA 2120
Page 1: ...ADVIA 2120 2120i Hematology Systems Operator s Guide 067D0157 01 Rev 201 0 ...
Page 30: ...1 26 Welcome to the ADVIA 2120 2120i Hematology System ...
Page 46: ...3 12 Daily Routine ...
Page 171: ...7 2 Status Line messages T 93 U 94 V 96 W 98 X 99 Y 99 Z 99 ...
Page 269: ...7 100 Status Line messages ...
Page 341: ...8 72 Methods ...