Status Line messages
7-73
PEROX Saturation - Alarm
The number of consecutive occurrences of this condition has met the criterion
specified for an alarm message. The specified alarm criterion is user definable in
the Customize Menu, System Setup, Alarm/Stop Criteria. The counter associated
with the Alarm Criteria is automatically reset.
PEROX Saturation - Stop
The number of consecutive occurrences of this condition has met the criterion
specified for a stop message. The autosampler halts at the same time the stop
message occurs. The specified stop criterion is user definable in the Customize
Menu, System Setup, Alarm/Stop Criteria. The counter associated with the Stop
Criteria is reset when the Start/Stop button is selected.
PEROX SHEATH Pump Error - Reset Required
PEROX SHEATH syringe pump not in home position, failed to respond, or a
transmission error occurred.
Possible Cause
1
PEROX SHEATH
syringe pump not
moving or not in
home position.
Corrective Action
a
Reset the analyzer by selecting
Utilities,
Exerciser, Indicators, Analyzer Reset
.
b
After about 1 minute, if the status line does not
display Ready to Run and the green ready-to-
aspirate light is still off, follow these steps to exit
the ADVIA 2120/2120i software, turn off the
analyzer and then restart.
i
Select
Shut Down ADVIA
at the Log On/Off
window.
ii
Select
Off
at the analyzer touchpad.
iii
Select
CTRL+ALT+DELETE
and log back
onto the system to restart the software.
iv
Restart the analyzer by selecting
On
at the
analyzer touchpad.
2
PEROX SHEATH
syringe incorrectly
assembled.
Follow the steps in Replacing the Syringe Plungers.
3
Faulty power
supply, fuse, cable,
pump assembly,
Can Bus Scrambler
Board, or Dual
Servo Pump
Control Board.
Call Siemens Service for assistance.
Summary of Contents for ADVIA 2120
Page 1: ...ADVIA 2120 2120i Hematology Systems Operator s Guide 067D0157 01 Rev 201 0 ...
Page 30: ...1 26 Welcome to the ADVIA 2120 2120i Hematology System ...
Page 46: ...3 12 Daily Routine ...
Page 171: ...7 2 Status Line messages T 93 U 94 V 96 W 98 X 99 Y 99 Z 99 ...
Page 269: ...7 100 Status Line messages ...
Page 341: ...8 72 Methods ...