
ADVIA Autoslide Slide Maker Stainer
10-5
Main power switch and power supply connection
CAUTION
The power switch and input voltage supply connection should always be accessible.
When positioning the system for operational use, leave the required amount of space for
easy accessibility to these items.
Installation, connection, and disconnection of the ADVIA Autoslide Slide Maker Stainer
must be performed by authorized personnel only.
For the main supply, grounding is required. Check that earth wall-plug is correctly
connected to the laboratory grounding system. If there is no such system, a ground stake
should be used.
Use only main supply cable delivered with the instrument or cable with same
characteristics:
•
Europ cord type: HO5VV-F
•
USA cord type: SVT,AWG 18-3
Main supply voltage fluctuations not to exceed
±
10% of the nominal voltage.
Fluidic and sensor connections
BIOHAZARD
Wear personal protective equipment. Use universal caution. For complete biohazard statement,
refer to section on
Autoslide Safety Information and Warnings
.
METHANOL
SAFETY
For complete methanol safety statement, refer to section on
Autoslide Safety Information and
Warnings
.
Summary of Contents for ADVIA 2120
Page 1: ...ADVIA 2120 2120i Hematology Systems Operator s Guide 067D0157 01 Rev 201 0 ...
Page 30: ...1 26 Welcome to the ADVIA 2120 2120i Hematology System ...
Page 46: ...3 12 Daily Routine ...
Page 171: ...7 2 Status Line messages T 93 U 94 V 96 W 98 X 99 Y 99 Z 99 ...
Page 269: ...7 100 Status Line messages ...
Page 341: ...8 72 Methods ...