
7-20
Status Line messages
3
Air lines
supplying the
centering
collar are
pinched.
a
Open the autosampler access door.
b
Check for kinked or pinched tubing. Replace if
necessary.
c
Close the autosampler access door. The autosampler
will reset.
d
If the problem persists, call Siemens Service for
assistance.
4
Faulty output-
queue sensor
or wiring.
Call Siemens Service for assistance.
Autosampler - Collar Failed to Retract, Reset Autosampler
The autosampler centering collar failed to retract properly.
Possible Cause
1
Critical
Sampler
Error. Partial
Reset is
required.
Corrective Action
a
Open the autosampler access door.
b
Close the autosampler access door. The autosampler
will reset.
2
Critical
Sampler
Error. Full
Reset is
required.
a
At the analyzer touchpad, select
Off
.
b
Wait about 1 minute and select
On
to restart the
analyzer.
c
If the problem persists, call Siemens Service for
assistance.
3
Air lines
supplying the
centering
collar are
pinched.
a
Open the autosampler access door.
b
Check for kinked or pinched tubing. Replace if
necessary.
c
Close the autosampler access door. The autosampler
resets.
d
If the problem persists, call Siemens Service for
assistance.
4
Faulty output-
queue sensor
or wiring.
Call Siemens Service for assistance.
Autosampler - Centering Collar Motion Denied, Reset Autosampler
The autosampler attempted an inappropriate centering collar motion. An
autosampler reset is required.
Summary of Contents for ADVIA 2120
Page 1: ...ADVIA 2120 2120i Hematology Systems Operator s Guide 067D0157 01 Rev 201 0 ...
Page 30: ...1 26 Welcome to the ADVIA 2120 2120i Hematology System ...
Page 46: ...3 12 Daily Routine ...
Page 171: ...7 2 Status Line messages T 93 U 94 V 96 W 98 X 99 Y 99 Z 99 ...
Page 269: ...7 100 Status Line messages ...
Page 341: ...8 72 Methods ...