3-6
Daily Routine
Select
Create
to make a new workorder for this patient.
Select
New Patient
to create a new patient file for this name.
4.
Enter sample and patient information.
5.
Use the
Tab
key to move from field to field.
You
must
enter a sample ID number.
6.
Select tests.
7.
Select
OK
to confirm the entries.
The Access dialog box displays for the next workorder.
Listing the Pending Workorders
1.
At the Customize menu, select
Tools View
.
2.
When the tools list appears, double select
File Management
.
3.
Select the
Pending
check box.
4.
Optional:
Use the To
and
From date boxes to restrict listing to workorders
created within a specific period of time.
5.
Optional:
Choose a report format from the Format list. DEFAULT is usually
sufficient.
6.
Select
Display
to view the list, or select
to print it.
Running the Samples
Control Recommendations
•
Run controls in accordance with your laboratory protocol.
•
Run multilevel controls at the beginning of each shift.
Siemens recommends the use of ADVIA TESTpoint Hematology Controls
(Low, Normal, and High) and ADVIA TESTpoint Reticulocyte Control
(Low and High).
•
Run controls alone or at the start of the patient workload.
•
Optionally, you can run a retained patient sample periodically to monitor
performance trends.
•
Evaluate all control results before reporting patient results.
•
If control results fail to meet the laboratory’s established criteria for
acceptability, you must evaluate all patient test results obtained in the
unacceptable run to determine if the patient results were adversely affected.
•
Perform and document appropriate corrective actions, which may include
recalibration and reassaying of patient samples, before reporting patient
results.
Summary of Contents for ADVIA 2120
Page 1: ...ADVIA 2120 2120i Hematology Systems Operator s Guide 067D0157 01 Rev 201 0 ...
Page 30: ...1 26 Welcome to the ADVIA 2120 2120i Hematology System ...
Page 46: ...3 12 Daily Routine ...
Page 171: ...7 2 Status Line messages T 93 U 94 V 96 W 98 X 99 Y 99 Z 99 ...
Page 269: ...7 100 Status Line messages ...
Page 341: ...8 72 Methods ...