8-32
Methods
Location of Unlysed RBCs on PEROX Cytogram
Unlysed RBCs
with no
peroxidase activity appear
along the left side of the
PEROX cytogram.
e Method
Calculating Parameters
Reported Parameters
Parameter Explanation
WBCP
RawWBC x (PeroxCalFactor
÷
[1-FracDT ])
%NEUT
(100 x Neutrophil Count)
÷
PHA Cells
#NEUT
(%NEUT
÷
100) x WBC
%LYMPH
([100 x Lymphocyte Count]
÷
PHA Cells) -
%BASO
#LYMPH
(%LYMPH
÷
100) x WBC
%MONO
(100 x Monocyte Count)
÷
PHA Cells
#MONO
(%MONO
÷
100) x WBC
%EOS
(100 x Eosinophil Count)
÷
PHA Cells
#EOS
(%EOS
÷
100) x WBC
%LUC
(100 x LUC Count)
÷
PHA Cells
#LUC
(%LUC
÷
100) x WBC
The system uses data from the perox channel and the baso channel to derive the
Nucleated Red Blood Cell (NRBC) count (%NRBC and #NRBC). If nRBCs are
present, the system corrects the WBC results to account for them.
The system can also display Peroxidase channel and Basophil/Lobularity channel
WBC counts and the associated differential counts that have not been corrected
for NRBCs, even if NRBC counts are reported. The uncorrected counts are
designated by a lower case "u".