7-78
Status Line messages
Printer Problem - Alarm
A printer problem has occurred. The number of consecutive occurrences of this
condition has met the criterion specified for an alarm message. The specified
alarm criterion is user definable in the Customize Menu, System Setup,
Alarm/Stop Criteria. The counter associated with the Alarm Criteria is
automatically reset.
Corrective Action
Verify printer connections and operation. Typical printer problems include:
•
Paper jams
•
Out of paper
•
Printer offline
•
Printer disconnected
Printer Problem - Stop
A printer problem has occurred. The number of consecutive occurrences of this
condition has met the criterion specified for a stop message. The specified stop
criterion is user definable in the Customize Menu, System Setup, Alarm/Stop
Criteria. The counter associated with the Stop Criteria is reset when the
Start/Stop button is selected.
Corrective Action
Verify printer connections and operation. Typical printer problems include:
•
Paper jams
•
Out of paper
•
Printer offline
•
Printer disconnected
Probe Clog
Conductivity detector did not detect a sample within twenty seconds of
aspiration.
Possible Cause
1
Clogged
sample
Corrective Action
Check for sample clots.
2
Insufficient
sample volume
Sample tube may have been removed before aspiration
was complete. Re-aspirate sample. If error persists clean
the sample shear valve.
Summary of Contents for ADVIA 2120
Page 1: ...ADVIA 2120 2120i Hematology Systems Operator s Guide 067D0157 01 Rev 201 0 ...
Page 30: ...1 26 Welcome to the ADVIA 2120 2120i Hematology System ...
Page 46: ...3 12 Daily Routine ...
Page 171: ...7 2 Status Line messages T 93 U 94 V 96 W 98 X 99 Y 99 Z 99 ...
Page 269: ...7 100 Status Line messages ...
Page 341: ...8 72 Methods ...