7-94
Status Line messages
Check the sample status and the contents of the file before attempting to
retransmit this sample. The problem may be an acknowledgement problem at the
host level. The communication line should be checked by Siemens Service
personnel.
Type Not Found
The type of workorder sent by the host is unknown to the Data Manager. No type
can be associated with the test request.
Corrective Action
Create this type in the Type Table Dictionary. Select Customize menu, System
Setup, Tools Modify, Type Table Dictionary.
Types Table Test Doesn't Exist
A test defined in the Types Table has been deleted in the Test Dictionary.
Corrective Action
As appropriate, delete the test from the Type Table, or reenter the test in the Test
Dictionary.
U
UFC Vacuum Out of Range
UFC Vacuum is out of range. The difference between the compressor vacuum
and UFC vacuum is greater than 5" Hg.
Possible Cause
1
Vacuum Setting
is out of range
Corrective Action
Check the system vacuum gauge and adjust the regulator
if necessary.
Setpoint: 20"Hg
±
1
2
Vacuum leak
a
Check for crimped or disconnected hydraulic or
pneumatic lines. Straighten or reconnect lines as
necessary.
b
Check the waste container to make sure that the
spigot is closed and the cap is on securely.
3
Vacuum pump
filter
(vacushield) is
clogged
Replace the vacuum pump filter.
4
Crimped waste
line
Check for a crimped waste line on the waste container.
Straighten or reconnect lines as necessary.
Summary of Contents for ADVIA 2120
Page 1: ...ADVIA 2120 2120i Hematology Systems Operator s Guide 067D0157 01 Rev 201 0 ...
Page 30: ...1 26 Welcome to the ADVIA 2120 2120i Hematology System ...
Page 46: ...3 12 Daily Routine ...
Page 171: ...7 2 Status Line messages T 93 U 94 V 96 W 98 X 99 Y 99 Z 99 ...
Page 269: ...7 100 Status Line messages ...
Page 341: ...8 72 Methods ...