10-44
ADVIA Autoslide Slide Maker Stainer
Analyzer mode: Standby
To replace the smearing tape
1.
At the Operations menu select
Autoslide Control
.
The Autoslide Control window opens.
2.
Under the Special Functions section, select
Smearing
Tape
.
3.
Under Smearing Tape Options, select
Replace
Tape
.
4.
Open the Autoslide lower front cover.
5.
Using the black magnetic knob, open the tape holder plate.
6.
Remove the upper and lower bobbins from their holders.
7.
Each time the smearing tape is replaced, wipe the wedge using a soft cloth to
remove any dust.
8.
Place the new smearing tape bobbin on the upper holder and the take up
bobbin on the lower holder. Be careful not to twist or crimp the tape. Thread
it through the spindles as shown in the diagram on the inside of the front
cover.