Methods
8-37
Perox Noise
The number of events in the
Noise area (1) of the
PEROX cytogram
.
Perox Saturation
The number of events in the
Saturation area (1) of the
PEROX cytogram.
RBC / Platelet Method
Cytochemical Reactions
The RBC/Platelet cytochemical reactions consist of two steps:
Step 1
RBCs and platelets are isovolumetrically sphered using the ADVIA
2120/2120i RBC/PLT reagent.
Step 2
RBCs and platelets are fixed.
ADVIA 2120/2120i RBC / PLT
NOTE:
For more detailed information on the contents of ADVIA 2120/2120i
RBC / PLT reagent, please see Chapter 8.
ADVIA 2120/2120i RBC/PLT reagent contains sodium dodecyl sulfate (SDS)
and glutaraldehyde, which causes sphering of the red blood cells and platelets.
When red cells and platelets are isovolumetrically sphered, shape is eliminated as
a variability factor.
Summary of Contents for ADVIA 2120
Page 1: ...ADVIA 2120 2120i Hematology Systems Operator s Guide 067D0157 01 Rev 201 0 ...
Page 30: ...1 26 Welcome to the ADVIA 2120 2120i Hematology System ...
Page 46: ...3 12 Daily Routine ...
Page 171: ...7 2 Status Line messages T 93 U 94 V 96 W 98 X 99 Y 99 Z 99 ...
Page 269: ...7 100 Status Line messages ...
Page 341: ...8 72 Methods ...