7-54
Status Line messages
i
Select
Shut Down ADVIA
at the Log
On/Off window.
ii
Select
Off
at the analyzer touchpad.
iii
Press
CTRL+ALT+DELETE
and log back
onto the system to restart the software.
iv
Restart the analyzer by selecting
On
at
the analyzer touchpad.
2
The cable connected to
the switch/indicator is
crimped.
Select
Off
at the analyzer touchpad to check for
crimped cable. Free the crimped cable if
necessary and restart the analyzer by selecting
On
at the analyzer touchpad.
3
Faulty power supply,
fuse, Switch/Indicator
board, cable, Parallel
Node Board, Can Bus
Scrambler Board, or
Switch/Indicator panel.
Call Siemens Service for assistance.
Initialization Failed - During Communication from Analyzer
Software communication error.
Corrective Action
Follow these steps to exit the ADVIA 2120/2120i software and then restart:
1
Select
Shut Down ADVIA
at the Log On/Off window.
2
Select
CTRL+ALT+DELETE
and log back onto the system to restart the
software.
If the problem persists, call Siemens Service for assistance.
Initialization Failed - During Configuration Download
Software communication error.
Corrective Action
Follow these steps to exit the ADVIA 2120/2120i software and then restart:
1
Select
Shut Down ADVIA
at the Log On/Off window.
2
Select
CTRL+ALT+DELETE
and log back onto the system to restart the
software.
If the problem persists, call Siemens Service for assistance.
Initialization Failed - During Download of Cycle Definition
Software communication error.
Summary of Contents for ADVIA 2120
Page 1: ...ADVIA 2120 2120i Hematology Systems Operator s Guide 067D0157 01 Rev 201 0 ...
Page 30: ...1 26 Welcome to the ADVIA 2120 2120i Hematology System ...
Page 46: ...3 12 Daily Routine ...
Page 171: ...7 2 Status Line messages T 93 U 94 V 96 W 98 X 99 Y 99 Z 99 ...
Page 269: ...7 100 Status Line messages ...
Page 341: ...8 72 Methods ...