7-30
Status Line messages
Error.
will reset.
c
If the problem persists, select
Off
at the analyzer
touchpad.
d
Wait about 1 minute and select
On
to restart the
analyzer.
e
If the problem persists, call Siemens Service for
assistance.
2
40 PSI line is
out of range.
a
Check pressure and adjust if necessary.
b
Check for pinched air lines supplying needle
cylinder.
c
At the analyzer touchpad, select
Off
.
d
Wait about 1 minute and select
On
to restart the
analyzer.
e
If the problem persists, call Siemens Service for
assistance.
3
Needle is
bent.
Replace needle.
4
Needle sensor
failed.
Call Siemens Service for assistance.
Autosampler - Needle Jammed
The needle failed to extend or retract during sample aspiration.
Possible Cause
1
40 PSI line out
of range.
Corrective Action
a
Check pressure and adjust if necessary.
b
Check for pinched air lines supplying needle
cylinder.
c
At the analyzer touchpad, select
Off
.
d
Wait about 1 minute and select
On
to restart the
analyzer.
e
If the problem persists, call Siemens Service for
assistance.
2
Needle sensor
failed.
Call Siemens Service for assistance.
Autosampler - Needle Motion Denied, Reset Autosampler
The autosampler attempted an inappropriate needle motion. An autosampler reset
is required.
Corrective Action
Summary of Contents for ADVIA 2120
Page 1: ...ADVIA 2120 2120i Hematology Systems Operator s Guide 067D0157 01 Rev 201 0 ...
Page 30: ...1 26 Welcome to the ADVIA 2120 2120i Hematology System ...
Page 46: ...3 12 Daily Routine ...
Page 171: ...7 2 Status Line messages T 93 U 94 V 96 W 98 X 99 Y 99 Z 99 ...
Page 269: ...7 100 Status Line messages ...
Page 341: ...8 72 Methods ...