Status Line messages
7-79
3
Insufficient
vacuum
Check vacuum gauge and adjust if necessary.
4
Clogged clot
filter
Remove clot filter and discard as biohazardous material.
Replace clot filter
.
5
Loose
connection in
sample line
Check connections and tighten if necessary
.
6
Clogged needle
Rinse needle. Select
Utilities
menu,
Hydraulic
Functions
tab,
Needle
Rinse
.
7
Bent or
damaged
needle
Replace the needle.
Follow the steps in Replacing the Sampler Needles
8
Faulty
Conductivity
Detector
Contact your local Siemens technical support provider
for assistance.
Q
QC Sample Deleted Before Validation
A control sample that was not validated has been deleted.
Daily QC may not be consistent.
R
Rack or Position Barcode Not Read - Alarm
The barcode reader cannot determine a valid rack ID or position number for the
current tube position. The number of consecutive occurrences of this condition
has met the criterion specified for an alarm message in the Alarm/Stop Criteria
window.
Possible Cause
1
Barcode label
improperly
placed,
obscured, or
absent.
Corrective Action
Check that a barcode label is properly attached and
meets specifications. Replace if necessary.
2
Faulty barcode
reader.
Call Siemens Service for assistance.
Summary of Contents for ADVIA 2120
Page 1: ...ADVIA 2120 2120i Hematology Systems Operator s Guide 067D0157 01 Rev 201 0 ...
Page 30: ...1 26 Welcome to the ADVIA 2120 2120i Hematology System ...
Page 46: ...3 12 Daily Routine ...
Page 171: ...7 2 Status Line messages T 93 U 94 V 96 W 98 X 99 Y 99 Z 99 ...
Page 269: ...7 100 Status Line messages ...
Page 341: ...8 72 Methods ...