Status Line messages
7-43
Autosampler
software error
1
At the analyzer touchpad, select
Off
.
2
Wait about 1 minute and select
On
to restart the
analyzer.
3
If problem persists, call Siemens Service for
assistance.
Check Waste Container Vacushield
The vacuum pump filter (vacushield) on the waste container may be saturated
with liquid.
Possible Cause
1
Vacuum pump
filter
(vacushield) is
clogged
Corrective Action
Replace the vacuum pump filter.
2
Vacuum leak
a
Check for crimped or disconnected hydraulic or
pneumatic lines. Straighten or reconnect lines as
necessary.
b
Check the waste container to make sure that the spigot
is closed and the cap is on securely.
3
Crimped waste
line
Check for a crimped waste line on the waste container.
Straighten or reconnect lines as necessary.
IMPORTANT
If problems persist, Call Siemens Service for assistance. Further corrective action
must be taken by Siemens service personnel only.
Communications Error with Analyzer
Software communication error between the computer and the analyzer.
Possible Cause
1
Internal
software
corruption
Corrective Action
a
Reset the analyzer by selecting
Utilities, Exerciser,
Indicators, Analyzer Reset
.
b
After about 1 minute, if the status line does not
display Ready to Run and the green ready-to-
aspirate light is still off, follow these steps to exit the
ADVIA 2120/2120i software, turn off the analyzer,
and then restart.
i
Select
Shut Down ADVIA
at the Log On/Off
window.
ii
Select
Off
at the analyzer touchpad.
iii
Select
CTRL+ALT+DELETE
and log back onto
the system to restart the software.
Summary of Contents for ADVIA 2120
Page 1: ...ADVIA 2120 2120i Hematology Systems Operator s Guide 067D0157 01 Rev 201 0 ...
Page 30: ...1 26 Welcome to the ADVIA 2120 2120i Hematology System ...
Page 46: ...3 12 Daily Routine ...
Page 171: ...7 2 Status Line messages T 93 U 94 V 96 W 98 X 99 Y 99 Z 99 ...
Page 269: ...7 100 Status Line messages ...
Page 341: ...8 72 Methods ...