Status Line messages
7-3
%
% Disk Space Used - Alarm
Disk space used has risen to the level specified as an alarm criterion, indicating
available disk space is low.
Corrective Action
Increase available disk space by deleting data from any of these files:
•
Raw Data files
•
Export files
•
Sample Results (use Customize, System Setup, Tools Modify, End of Day)
•
Logs: Message Log, Service Log, User Notepad, Reagent Log, Workload
Log
IMPORTANT
Deleting files permanently erases data. If you need an electronic copy of the data,
back up or export the files before deleting.
% Disk Space Used - Stop
Disk space used has risen to the level specified as the stop criterion, and the
autosampler has halted. If stop/alarm criteria are not configured, the default level
value is set to 98%.
Corrective Action
Increase available disk space by deleting data from any of these files:
•
Raw Data files
•
Export files
•
Sample Results (use Customize, System Setup, Tools Modify, End of Day)
•
Logs: Message Log, Service Log, User Notepad, Reagent Log, Workload
Log
IMPORTANT
Deleting files permanently erases data. If you need an electronic copy of the data,
back up or export the files before deleting.